Optimized for low bunch charge and low current
Noise in CW beam and macropulse measurement 0.1 µArms | Noise in single bunch measurement 10 fC |
80dB measurement range without range switching
Resolution 1% / accuracy 4% of measured value
DC Output voltage, logarithmically proportional to input current or input charge
USB 2.0 controls and readout up to 5kS/s
Negligible magnetic field dependence
UHV compatible down to 1E-10 mbar
Two packaging types for the Turbo-ICT sensor
Turbo-ICT is mounted directly on the beam line |
Vacuum compatible to 1E-7 mbar Calibrated charge generator option not available 1 core or 2 cores on option |
BCM-RF-E allows two modes of operation:
Output is logarithmically proportional to beam current or bunch charge. Up to 10 stations per chassis can be installed. |
CW beam and macropulses | Single bunch | |
Measurement range | 0.5µA – 3mA | 50fC – 300pC |
Bunch repetition | 10MHz – 350MHz | Single bunch <2MHz |
Output | 0 - +5V, log of | 0 - +5V, log of |
Noise | 0.1µArms | 10fC |